# AIGxG User Manual UHV Bayard-Alpert Ion Gauge # Installation Inspect the gauge structure before removing from the moulded clear plastic packing. Report any obvious damage to your AML Distributor before opening the moulded packing. Save the packing material, including the small foam plastic piece which restrains the flange in the mould. It is assumed that the user is familiar with the use of ConflatĀ® flanges and copper gaskets. Appropriate handling techniques must be used to avoid contamination of the gauge head. Avoid straining the feedthrough pins, as this may cause a leak. Mount the gauge head in a position where the free electrons generated in its vicinity will not affect other equipment. The gauge head is intended to be inserted so that the entire grid structure is inside the chamber: should any part of it remain inside the tube behind the mounting flange the sensitivity of the gauge head will be altered. The orientation of the gauge head should be such that the filaments are to the side of, or below, the grid structure. This will ensure that if the filament should sag or break it will not short-circuit to the grid. It is desirable that the feedthrough pins point sideways or upward so that the weight of the connector and cable do not tend to pull the connector off the pins.
[![AIG1.png](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/scaled-1680-/aig1.png)](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig1.png)[![AIG2.png](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/scaled-1680-/aig2.png)](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig2.png) [![AIG3.png](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/scaled-1680-/aig3.png)](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig3.png)[![AIG4.png](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/scaled-1680-/aig4.png)](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig4.png)
NoYesYesNo
The performance of the ion gauge may be affected by other electron or ion generating processes within the vacuum chamber: should shielding of the gauge head be necessary, ensure that the conductance between the gauge head and volume of interest is not significantly decreased by its presence. Any grounded shield will affect the electron dynamics in the gauge head and hence its sensitivity: this can be minimised by locating the shield as far away as is convenient. When this gauge is used in conjunction with an AML NGC3 Ion Gauge Controller, both the gauge and controller are protected from all normal failure modes of either. Users should be aware of potential hazards from other equipment, however, particularly those introducing high voltages into the vacuum chamber (X ray sources for example). As a direct discharge from one of these at high pressure may cause extensive damage, shielding should always be introduced in such cases. # Ion Gauge Cables

**WARNING!** The use of AML AIGLx cables is essential to achieve compliance with the EC EMC Directive, 89/336/EEC using AML and similar controllers. Extension cables are not recommended as these will prejudice compliance. 3, 6 and 9 metre cables are available, models AIGL3, AIGL6 and AIGL9, respectively.

[![AIG Cable1.png](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/scaled-1680-/aig-cable1.png)](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig-cable1.png)The composite cables are heavy and require to be strain-relieved near both ends. This can be achieved by supporting the cable with ties at points 's' shown in the diagrams. Distance 'd' should be between 20 and 50cm. The cable should be slack between the tie and the connector. In bakeout zones stainless-steel cable ties should be used.
[![AIG Cable2.png](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/scaled-1680-/aig-cable2.png)](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig-cable2.png)At the instrument end of the cable the connector pins 'float' inside the moulding. Always support and align the cable at the rear of the connector when inserting it into the controller connector as a misaligned pin may be damaged if forced. The green and yellow flying ground lead must be connected to the M5 ground stud on the rear panel of the NGC3.
# Ion Gauge Operation New gauges must be operated carefully to achieve good results and long filament life. The filaments are under tension when new: run both filaments for a few seconds in emission to reduce and equalise the tension. Operate at low emission current in order to outgas them fully before increasing the emission current. Do not subject the gauge to electron- bombardment degas until it has run at the highest emission current available and the resultant pressure burst has fully subsided. The signal current from a BA gauge at UHV is very small and can easily be disturbed. Do not allow the cable to be moved in use as this will induce currents and generate piezoelectric and triboelectric charges which can affect UHV measurements for several seconds. Ensure that the front face of the AIGLx ion gauge connector is kept clean, since effective insulation resistances of the order of 1016 ohms are required. Isopropyl alcohol or CFC on a clean lint-free swab may be used to remove fingerprints or dirt. These solvents do not affect the PEEK housing of the connector. The nominal bias conditions and sensitivity for this gauge are given in the table below. AML and similar controllers provide these bias voltages. ### Recommended operating conditions
Emission Degas
Collector +0V +0V
Grid +200V +500V
Filament bias +50V +0V
Maximum Emission Current 10mA 100mA Tungsten 60mA Thoriated
# Filament Replacement Tungsten filaments have an emission life of one to three years in UHV. The life will be dependent on the residual gases present, range of pressures and vibration coupled to the filament. Tungsten filaments become brittle in use and the most common causes of failure are mechanical or thermal shocks. If the gauge head is to be moved to another location it is advisable to replace the filament assembly. Replacement filament assemblies are available from AML distributors at low cost. The filament assemblies are fixed to the feedthrough assembly by barrel connectors and Allen-headed set screws. An Allen key is provided with each replacement filament assembly. A pair of long-nosed pliers, side cutters, cleanroom gloves, cotton buds and some isopropyl alcohol are required.
[![AIG FIL Assy.png](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/scaled-1680-/aig-fil-assy.png)](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig-fil-assy.png)Clean the jaws of the cutters and pliers with alcohol. Swab some alcohol onto the set screws to lubricate them. Support the feedthrough wire with the long-nosed pliers close to the barrel connector and loosen the screw holding the filament support. This is the screw furthest away from the feedthrough flange. Loosen the screws on all three supports and remove the old filament assembly. Remove the restraining sleeve from the new assembly. Fit the centre support of the new assembly into the barrel connector and tighten the screws. Hold each of the filament terminations in turn with the pliers and extend them down to the outer barrel connectors. Fix them in place with the set screws. Ensure that the centre support of the filament assembly is parallel to the gauge axis and approximately 3mm from the side of the grid.
# Repair and Refurbishment Other than gauge heads returned under warranty it is generally not economically feasible to return damaged gauges for repair. Leaking feedthroughs cannot be repaired. Distorted grid structures and failed filaments are readily repairable by users. Any gauge head returned to AML must be accompanied by a signed statement by the user stating clearly the nature of any toxic or radioactive material used in its operating environment and the means taken to ensure any contamination is removed. Ensure that gauge heads are properly packed in original AML packing. # Gas Correction Factors Gas correction factor tables are are only reproduced for the convenience of the user and do not imply that use with other gases will be safe with BA ion gauges. Divide sensitivity by 100 for Pa-1; multiply by 1.33 for Torr-1.
Gas Symbol Gas Correction Factor NGC Sensitivity S, mBar-1
Acetone (CH3)2CO 3.6 68
Air --- 1.0 19
Ammonia NH3 1.3 25
Argon Ar 1.3 24
Benzene C6H6 5.9 112
Bromine Br 3.8 72
Bromomethane CH3Br 3.7 70
Cadmium Cd 2.3 44
Carbon Dioxide CO2 1.4 27
Carbon Disulfide CS2 5.0 95
Carbon Monoxide CO 1.05 20
Carbon Tetrachloride CCl4 6.0 114
Cesium Cs 4.3 82
Chlorine CI2 0.68 13
Chlorobenzene C6H5CI 7.0 133
Chloroethane C2H5CI 4.0 76
Chloroform CHCI3 4.7 89
Chloromethane CH3CI 2.6 49
Cyanogen (CN)2 2.8 53
Cyclohexylene C6H12 7.9 150
Deuterium D2 0.35 7
Dichlorodifloromethane CCI2F2 2.7 51
Dichloromethane CH2CI2 3.7 70
Ethane C2H6 2.6 49
Ethanol C2H5OH 3.6 68
Ethyl Acetate CH3COOC2H5 5.0 95
Ethyl ether (C2H5)2O 5.1 97
Ethylene C2H4 2.3 44
Ethylene oxide (CH2)2O 2.5 47
Helium He 0.18 3
Heptane C7H16 8.6 163
Hexane C6H14 6.6 125
Hydrogen H2 0.46 9
Hydrogen Bromide HBr 2.0 38
Hydrogen Chloride HCl 1.5 28
Hydrogen Cyanide HCN 1.5 28
Hydrogen Fluoride HF 1.4 27
Hydrogen Iodide HI 3.1 59
Hydrogen Sulfide H2S 2.2 42
Iodine I2 5.4 103
Iodomethane CH3I 4.2 80
Isoamyl Alcohol C5H11OH 2.9 55
Isobutylene C4H8 3.6 68
Krypton Kr 1.9 36
Lithium Li 1.9 36
Mercury Hg 3.6 68
Methane CH4 1.4 27
Methanol CH3OH 1.8 34
Methtyl Acetate CH3COOCH3 4.0 76
Mythyl ether (CH3)2O 3.0 57
Naphthalene C10H8 9.7 184
Neon Ne 0.3 6
Nitrobenzene C6H5NO2 7.2 137
Nitric Oxide NO 1.3 25
Nitrogen N2 1.0 19
Nitrogen Oxide NO2 1.2 23
Nitrous Oxide N2O 1.5 28
Oxygen O2 1.0 19
Phosphine PH3 2.6 49
Potassium K 3.6 68
Propane C3H8 4.2 80
Rubidium Rb 4.3 82
Sodium Na 3.0 57
Sulphur Dioxide SO2 2.1 40
Sulphur Hexafluoride SF6 2.3 44
Toluene C6H5CH3 6.8 129
Water H2O 1.1 21
Xenon Xe 2.9 55
# Assistance In the first instance, contact the distributor or supplier of the equipment. Always quote the serial number of the instrument and firmware and software versions. Provide a written description of the problem. If the problem is related to a motor or mechanism manufactured by AML, include the serial number(s) of those items. Do not return products to AML without prior approval. Arun Microelectronics Ltd Tel: +44 (0)1903 884141 Email: info@arunmicro.com Website: [arunmicro.com](https://arunmicro.com/)