# Installation Inspect the gauge structure before removing from the moulded clear plastic packing. Report any obvious damage to your AML Distributor before opening the moulded packing. Save the packing material, including the small foam plastic piece which restrains the flange in the mould. It is assumed that the user is familiar with the use of ConflatĀ® flanges and copper gaskets. Appropriate handling techniques must be used to avoid contamination of the gauge head. Avoid straining the feedthrough pins, as this may cause a leak. Mount the gauge head in a position where the free electrons generated in its vicinity will not affect other equipment. The gauge head is intended to be inserted so that the entire grid structure is inside the chamber: should any part of it remain inside the tube behind the mounting flange the sensitivity of the gauge head will be altered. The orientation of the gauge head should be such that the filaments are to the side of, or below, the grid structure. This will ensure that if the filament should sag or break it will not short-circuit to the grid. It is desirable that the feedthrough pins point sideways or upward so that the weight of the connector and cable do not tend to pull the connector off the pins.
[](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig1.png) | [](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig2.png) | [](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig3.png) | [](https://bookstack.vps-da8d40f3.arunmicro.com/uploads/images/gallery/2024-01/aig4.png) |
No | Yes | Yes | No |